You signed in with another tab or window. Reload to refresh your session.You signed out in another tab or window. Reload to refresh your session.You switched accounts on another tab or window. Reload to refresh your session.Dismiss alert
The Lakeshore 336 is a temperature controller that has four input temperature readings and two heater outputs. The user can configure which heater output is paired up with which input.
IOC Setup
When connecting to the real device, the IOC requires its IP address, but the software installed on the device has a bug and does not return the correct IP address. Its network name, however, is correct. To use the device, set the "IPADDR" macro oin globals.txt to be the device network name, e.g. LKSH336_01__IPADDR=ls336-1.